Magnum® Pilot Freeze Dryer/Lyophilizer


Shelf area up to 20 sq ft (1.86 sq M) & 30 L condenser





APPLICATIONS

  • Tissue Bank
  • Small Scale Production
  • Diagnostics
  • Well Plates
  • Vials
  • Bulk Applications / Trays





Offering up to 20 sq ft of shelf area and a condensing rate of 30 liters, the Magnum Pilot Freeze Dryer stands out amongst the competition.
This lyophilizer offers 30% more shelf area than most other units in its class. While the Magnum is commonly used for R&D purposes in the lab, it is the perfect lyophilizer for small scale production.




Our easy-to-use software provides automatic freeze-drying, defrost, system and leak rate testing.
Data can be printed either numerically or graphically. This same control system is used on industrial dryers, allowing scaling to production. An interactive maintenance screen simplifies component servicing.
All systems are remotely accessible, with customer approval, for troubleshooting process issues.




SYSTEM PERFORMANCE



  • 10 shelf pull down from +20 to –40 °C in less than 40 minutes *
  • Vacuum pull down to 100 mT in less than 20 minutes *
  • Vacuum leak rate less than 30 mT per hour *
  • Vacuum level 10 mT in clean dry system (-85 °C)

MAGNUM® STANDARD SPECIFICATIONS

SHELF AREA10 to 20 sq ft ( 0.929 to 1.858 sqM)
SHELF ASSEMBLY Bulk or Hydraulic Stoppering
SHELF TEMPERATURE RANGE -45 °C or –70 °C to +65 °C
SHELF HEAT TRANSFER Hollow Fluid Filled
SHELF SIZE/FINISH12"x 24”, 316LSS, 20 Ra or better (304 x 304 mm)
CONDENSER TEMPERATURE -53 °C or –85 °C
CONDENSER CAPACITY 30 L
CONDENSING RATE 20 L in 24 hours
CONDENSER STYLE Exposed Coil, 6” port
DEFROST Hot Gas
COMPRESSORS (SCROLL) 5 HP (-53 °C), 3.5 HP & 2 HP(-85 °C)
PRODUCT SENSORS 4, Type T Thermocouples
VACUUM PUMP 375 LPM,
VACUUM CONTROL Pirani w/ Solenoid & Needle Valve
GAS BACKFILL Included
CONTROL SYSTEM PC/PLC with Opti-Dry® Software
TRAYS One per shelf included
CABINET 35" w x 46"d x 75"h
ELECTRICAL FOR -53 °C UNITS230V/60Hz, 3 ph, 30 A | 460V/60Hz, 3ph, 25 A | 380V/50Hz, 3 ph, 30A | 220V/50Hz, 3 ph, 30 A
ELECTRICAL FOR -85 °C UNITS230V/60Hz, 1 ph, 40 A | 220V/50Hz, 1 ph, 40 A | 380V/50Hz, 3 ph, 30A
PLEASE NOTE THAT 50HZ UNITS HAVE HEAT REMOVAL RERATED BY 17%